In order to study the effects of active speed modulation on the internal flow field and blood damages in the blood pumps, the CFD method was adopted to simulate the full flowpath internal flow of a blood pump under speed modulations. The combined numerical simulation of the lumped cardiovascular system mathematical model and the rotary blood pump model was used to obtain the ventricular and aortic pressures under the assisting of the blood pump, which were then set as the inlet and outlet boundary conditions of the blood pump in CFD simulations. The flow fields of blood pump under the constant speed and three types of asynchronous speed modulation waveforms, including sine, square and triangle waves, were analyzed, and the velocity distribution and shear stress distribution of the rotary blood pumps were obtained. The results show that the flow pulsation of blood pump is enhanced under speed modulations, which is a feasible scheme to restore the pulsation of the blood flow. Among the three speed modulation waveforms, the blood pump flow pulsation index is high and the shear stress in the blood pump is small under the sinusoidal speed modulation, which is a relatively ideal speed modulation waveform.
HUANG Feng, WANG Chuchen, RUAN Xiaodong
. Flow Field Analysis of a Centrifugal Rotary Blood Pump under Active Speed Modulations[J]. China Mechanical Engineering, 2021
, 32(24)
: 2915
-2923
.
DOI: 10.3969/j.issn.1004-132X.2021.24.003
[1]陈伟伟,高润霖,刘力生,等. 中国心血管病报告2017概要 [J]. 中国循环杂志,2018,33(1):1-8.
CHEN Weiwei, GAO Runlin, LIU Lisheng, et al. Chinese Cardiovascular Disease Report 2017 Summary [J]. Chinese Journal of Circulation, 2018, 33(1):1-8.
[2]李寰. 心室辅助装置溶血性能评价方法及叶轮优化设计研究 [D]. 杭州:浙江大学,2018.
LI Huan. Research on the Hemolytic Performance Evaluation and Impeller Optimization Design of VADs [D].Hangzhou:Zhejiang University, 2018.
[3]武悦,朱良凡,罗云. 计算流体力学方法分析一例喷射悬浮血泵的液力、悬浮及溶血特性 [J]. 机械工程学报,2018,54(20):52-58.
WU Yue, ZHU Liangfan, LUO Yun. Computational Fluid Dynamics Analysis of an Injection Suspension Blood Pump on the Hydraulic, Suspension and Hemolysis Property [J]. Journal of Mechanical Engineering, 2018, 54(20):52-58.
[4]刘彤,宋德利,刘小慧,等. 机械循环辅助在心力衰竭中的应用 [J]. 心肺血管病杂志,2015, 34(10):782-787.
LIU Tong, SONG Deli, LIU Xiaohui, et al. Application of Mechanical Circulatory Assistance in Heart Failure [J]. Journal of Cardiovascular and Pulmonary Diseases, 2015, 34(10):782-787.
[5]GRAY N A, SELZMAN C H. Current Status of the Total Artificial Heart [J]. American Heart Journal, 2006, 152(1):4-10.
[6]LOOR G, GONZALEZ-STAWINSKI G. Pulsatile vs. Continuous Flow in Ventricular Assist Device Therapy [J]. Best Practice & Research Clinical Anaesthesiology, 2012, 26(2):105-115.
[7]KRABATSCH T, SCHWEIGER M, DANDEL M, et al. Is Bridge to Recovery More Likely with Pulsatile Left Ventricular Assist Devices than with Nonpulsatile-flow Systems[J]. The Annals of Thoracic Surgery, 2011, 91(5):1335-1340.
[8]HUANG F, RUAN X, FU X. Pulse-pressure-enhancing Controller for Better Physiologic Perfusion of Rotary Blood Pumps Based on Speed Modulation [J]. ASAIO Journal, 2014, 60(3):269-279.
[9]HUANG F, GOU Z, FU Y, et al. Effects on the Pulmonary Hemodynamics and Gas Exchange with a Speed Modulated Right Ventricular Assist Rotary Blood Pump:a Numerical Study [J]. BioMedical Engineering OnLine, 2018, 17(1):No. 142.
[10]ISING M, WARREN S, SOBIESKI M, et al. Flow Modulation Algorithms for Continuous Flow Left Ventricular Assist Devices to Increase Vascular Pulsatility:a Computer Simulation Study [J]. Cardiovascular Engineering and Technology, 2011, 2:90-100.
[11]SOUCY K G, GIRIDHARAN G A, CHOI Y, et al. Rotary Pump Speed Modulation for Generating Pulsatile Flow and Phasic Left Ventricular Volume Unloading in a Bovine Model of Chronic Ischemic Heart Failure [J]. Heart Lung Transplant, 2015, 34(1):122-131.
[12]ZIMPFER D, STRUEBER M, AIGNER P, et al. Evaluation of the Heart Ware Ventricular Assist Device Lavare Cycle in a Particle Image Velocimetry Model and in Clinical Practice [J]. European Journal of Cardio-thoracic Surgery, 2016, 50:839-848.
[13]沈朋. 离心式血泵内部流场数值模拟及血损研究[D]. 哈尔滨:哈尔滨理工大学,2017.
SHEN Peng. Numerical Simulation of Internal Flow Field and Blood Damage Study of Centrifugal Blood Pump [D]. Harbin:Harbin University of Science and Technology, 2017.
[14]朱卓玲,赵伟国,黄峰. 基于BP神经网络的旋转血泵生理控制 [J]. 中国生物医学工程学报,2019,38(5):581-589.
ZHU Zhuoling, ZHAO Weiguo, HUANG Feng. Physiological Control of Rotary Blood Pumps Based on BP Neural Network [J]. Chinese Journal of Biomedical Engineering, 2019, 38(5):581-589.
[15]LI H, GOU Z, HUANG F, et al. Evaluation of the Hemolysis and Fluid Dynamics of a Ventricular Assist Device under the Pulsatile Flow Condition[J]. Journal of Hydrodynamics, 2019, 31(5):965-975.
[16]BLUDSZUWEIT C. Model for a General Mechanical Blood Damage Prediction [J]. Artificial Organs, 1995, 19:583-598.
[17]CHEN Z, MONDAL N K, DING J, et al. Paradoxical Effect of Nonphysiological Shear Stress on Platelets and von Willebrand Factor [J]. Artificial Organs, 2016, 40(7):659-668.
[18]HASIN T, DEO S, MALESZEWSKI J J, et al. The Role of Medical Management for Acute Intravascular Hemolysis in Patients Supported on Axial Flow LVAD [J]. ASAIO Journal, 2014, 60(1):9-14.
[19]TOLPEN S, JANMAAT J, REIDER C, et al. Programmed Speed Reduction Enables Aortic Valve Opening and Increased Pulsatility in the LVAD-assisted Heart [J]. ASAIO Journal, 2015, 61(5):540-547.